Resistance of medical gloves to permeation by Carmustine  (2001)

Authors

All employed at SMTL.

Introduction

The SMTL were requested to examine the permeability of medical gloves to the cytotoxic drug Carmustine. This document includes data generated for:

  1. Welsh Health Supplies: Test results of the latex examination and surgical gloves currently on the All Wales Contract for 2000-2001.
  2. Regional Pharmaceutical Committee: Test results of the nitrile examination glove currently on the All Wales Contract for 2000-2001.

Test Method and Definitions

Testing of the gloves was in accordance with BS EN 374 Part 3: 1994, Protective gloves against chemical and micro-organisms; Determination of resistance to permeation by chemicals. The principle of the test is to determine the resistance of a glove material to permeation by a solid or liquid chemical and for the purpose of this report, the alkylating drug Carmustine was utilised at a concentration of 100mg/30ml. With reference to the test results the following definitions apply:

Test results

Medisavers Bodyguards nitrile powderfree non sterile examination gloves, Medium

Sample No Sample location Thickness (mm) Breakthrough time (mins) Duration of test (mins) Maximum permeation rate (µg/min/cm2) Drug transfer (µg/cm2)
1 Palm 0.13 * 240[1] 0.37 23.8
2 Palm 0.15 * 240[1] 0.06 3.4
3 Palm 0.13 * 240[1] 0.03 1.1

Medisavers Bodyguards latex powderfree non sterile examination gloves, Medium

Sample No Sample location Thickness (mm) Breakthrough time (mins) Maximum permeation rate (µg/min/cm2) Drug transfer (µg/cm2)
1 Palm 0.18 31 1.13 15.7
2 Palm 0.17 25 1.19 9.7
3 Palm 0.17 21 1.30 9.8

LRC Biogel latex powderfree sterile surgical gloves, 7

Sample No Sample location Thickness (mm) Breakthrough time (mins) Maximum permeation rate (µg/min/cm2) Drug transfer (µg/cm2)
1 Palm 0.27 102 1.30 50.2
2 Palm 0.26 96 1.08 45.5
3 Palm 0.25 97 2.15 27.9

Semperit Sempermed Supreme latex powderfree sterile surgical gloves, 7

Sample No Sample location Thickness (mm) Breakthrough time (mins) Maximum permeation rate (µg/min/cm2) Drug transfer (µg/cm2)
1 Palm 0.22 46 1.32 16.2
2 Palm 0.20 41 1.47 17.6
3 Palm 0.21 45 1.42 16.1

Notes:

The maximum permeation rate and drug transfer are not specified in the standard. They are included for information only, especially when breakthrough of the drug does not occur.

0.0.0.1 * Where breakthrough did not occur, the duration of the test is reported as required by the standard.

0.0.0.2 NA denotes not applicable

 

breakthroughtime

Discussion

As can be seen from the tables above, the nitrile gloves did not show breakthrough (see above for definition) of the carmustine over the test period of 3 hours. The maximum permeation rate during the test was 0.37 µg/min/cm2 in one sample, the other two samples showing 0.06 and 0.03 µg/min/cm2 respectively.

The latex gloves, in contrast, all broke through well inside the 4 hour test period, with the examination gloves breaking through in around 25 minutes, and the two surgeons gloves at around 45 (Semperit) and 100 (LRC) minutes respectively.

Permeation rates for the latex gloves ranged between 1.1 and 2.15 µg/min/cm2.

The total amount of drug transferred (in the column headed `Drug Transfer') should be interpreted carefully, as the test is stopped once breakthrough is detected. This means that the test time for the Bodyguard latex glove was 20-30 minutes, and the total drug transfer of 10-15 µg/cm2 took place over this relatively short period, whilst the worst performing Bodyguard Nitrile glove transferred approximately 24µg/cm2 over a period of 4 hours.